Insertion of a pressure sensing array minimally affects hindfoot bone kinematics

Abstract

Understanding the development of ankle osteoarthritis (OA) is of high importance and interest; however its causality is poorly understood and several links to joint loading conditions have been made. One way of quantifying joint loading conditions is by measuring the intra-articular pressure distribution during gait simulations performed by in-vitro experimental set-ups. However the effect of inserting a pressure sensing array in the ankle joint could potentially disturb the proper kinematics and therefore the loading conditions. In this study, we performed in-vitro gait simulations in 7 cadaveric feet, before and after inserting a pressure sensing array and quantified the effect on the joints range of motion (ROM). The gait was simulated with a stance phase duration of one second using a custom build cadaveric gait simulator (CGS). The results show a limited effect in the ROM for all the joints of the hind foot, not exceeding the variability observed in specimens without a sensor. However, no consistent direction (increase/decrease) can be observed. The results suggest that even though the effect of inserting a pressure sensing array is minimal, it needs to be evaluated against the demands/requirements of the application.

Publication
In Journal of Foot and Ankle Research, Biomed Central